Auto Power off
High accuracy and rapid response
Low battery indication
Easy to use
Spesifications :
Power Supply : 9V battery (ANSI/NEDA 1604A or IEC 6LR619V alkaline)
Operating Temperature : 0°C ~ 50°C
Operating Relative Humidity : ≤80% RH
Storage Temperature : ≤-10°C ~ 55°C
Storage Relative Humidity : ≤90% RH
Size : 200x100x40mm (with holster)
Weight: 550g (with holster)
Standard Accessories : manual, industrial test lead (with alligator clips attached to probes)
Optional Accessories : AC power-supply adapter
| 
Function | 
Reference | 
Range | 
Resolution | 
Accuracy | 
Remark | 
| DC Voltage | 100mV | -10.00 ~ 110.00mV | 0.01mV | ±0.05% of set value or ±30µV | Maximum ouput current : ±2mA | 
| 1000mV | -100.00 ~ 1100.00mV | 0.1mV | ±0.05% of set value or ±0.3mV | ||
| OHM | 400Ω | 0.0 ~ 400Ω | 0.1Ω | ±0.05% of set value or ±0.2Ω | ±1mA excite current | 
| TC | R | -40°C ~ 1760°C | 1°C | ±0.05% of set value or ±3°C (≤100°C) ±0.05% of set value or ±2°C (>100°C) | By Using ITS-90 temperature scale | 
| S | -20°C ~ 1760°C | ||||
| B | 400°C ~ 1800°C | 1°C | ±0.05% of set value or ±3°C (400°C ~ 600°C) ±0.05% of set value or ±2°C (>600°C) | ||
| E | -200.0°C ~ 1000.0°C | 0.1°C | ±0.05% of set value or ±2°C (≤-100°C) ±0.05% of set value or ±1°C (>-100°C) | ||
| K | -200.0°C ~ 1370.0°C | 0.1°C | |||
| J | -200.0°C ~ 1200.0 °C | 0.1°C | |||
| T | -200.0°C ~ 400.0 °C | 0.1°C | |||
| N | -200.0°C ~ 1300°C | 0.1°C | |||
| RTD | Pt100-385 | -200.0°C ~ 850.0°C | 0.1°C | ±0.05% of set value or ±0.6°C | By using temperature scale Pt100-385. | 
| Pt200-385 | -50.0°C ~ 150 °C | 0.1°C | 
 
 
 
 
 
 

 




0 komentar: